Description
Into the varied landscape of HPLC autosamplers, the Prominence Series autosampler enters built on a solid foundation. Following the principle of Kaizen*, the SIL-20 incorporates and improves on many of the features that made its predecessors leaders in the mar- ket. Use of unique coatings and materials in critical areas, state-of-the-art electromechanical components, and innovative control through a Web browser ensure the SIL-20 will continue to set the standard that others will strive to match. Whatever you require – near-zero carryover, fast cycle times, or exceptional reproduc- ibility – the Prominence autosamplers will meet your need.
Shimadzu SIL-20 AC
| Injection Method |
Metering pump, variable volume, no sample loss |
| Injection Volume |
0.1 to 100 μL (std), 1 to 2,000 μL (opt) |
| Sample Capacity |
70 1.5 mL vials (incl. std.);
Control Rack: 10 1.5 mL vials
|
| Optional Sample Racks for: |
175 1 mL vials 50 4 mL vials 2 96 well MTP (std or deep well)
2 384 well MTP
|
| Injection Volume Accuracy |
+ 1% (50μL, n=10) |
| Injection Volume Precision |
RSD < 0.3% (10μL) |
| Carryover |
0.005% (Chlorhexidine) |
| Sample Aspiration Rate |
0.1-15 μL/sec (0.1μL steps) |
| Rinsing Rate |
1-35 μL/sec (1μL steps) |
| Rinsing Options |
Before Aspiration, After Aspiration, Before and After Aspiration, or None (Optional rinse pump for multiple solvent wash available) |
| Temperature Control |
4ºC-40ºC w/ built-in dehumidifier |
| pH Range |
1 to 14 |
| Operating Temperature |
4ºC to 35ºC (non-condensing atmosphere) |
| Dimensions & Weight |
260W x 415H x 500D, 30kg |
| Power Requirements |
AC110-120V, 300VA, 50/60Hz |